F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
J

Failure to Provide Safe Tracheostomy and Respiratory Care Resulting in Resident Deaths

Provo Rehabilitation And NursingProvo, Utah Survey Completed on 03-16-2026

Summary

The deficiency involves the facility’s failure to provide safe and appropriate respiratory and tracheostomy care in accordance with physician orders, professional standards, and resident-specific limitations. For one ventilator-dependent resident with chronic respiratory failure, tracheostomy status, pneumonia, anoxic brain damage, and documented subglottic and proximal tracheal stenosis, hospital records indicated that ENT had changed the trach in the OR and recommended trach changes every three months instead of monthly, and advised against trialing a Passy Muir Valve. A nurse practitioner note directed continued trach care per RT protocol with ENT follow-up for the stenosis. The DON reported seeing a provider note indicating the trach should be changed every three months, but this order was never entered into the medical record. The prior standing order to change the trach every 45 days had been converted to every 30 days by the contracted respiratory company, and then the trach change order dropped off the MAR in January, leaving no active order for trach changes at the time of the resident’s death. On the day of the event, RT documentation earlier in the day described the resident as alert and oriented, on a ventilator with an XLT cuffed, non-fenestrated size 7 trach, with the trach midline, secure, and patent, and no signs of respiratory distress. Later, an RT note documented that the RT performed a trach change because the resident reported the trach was too tight that morning. The RT pre-oxygenated the resident and recorded stable SpO2 and HR before the attempt. During removal and reinsertion of the trach tube, resistance was encountered and insertion was unsuccessful; a smaller trach was then attempted and also could not be inserted. The resident developed acute respiratory distress with pallor and cyanosis, oxygen saturation dropped, and a code blue was initiated. Nursing documentation corroborated that during the attempted trach change, resistance was met twice, the resident became cyanotic with decreasing oxygen saturation, and CPR was initiated after the resident was found unresponsive, apneic, and pulseless. EMS continued resuscitative efforts, and the resident was pronounced dead. Interviews revealed that RT 1 believed the resident’s complaint of tightness meant the trach needed to be changed immediately, pushed a partially dislodged trach back in earlier that morning, and then proceeded with a trach change despite the resident not being in distress and the trach being described as secure, patent, and midline. RT 2 stated he was unaware the trach was not supposed to be changed and had never changed a vented patient’s trach before. RT 3 stated there were instructions that the resident’s trach was not to be changed until seen by ENT, and that he had seen the resident’s name on a whiteboard with an “ENT only” notation. The RT Director stated she had received a verbal order that the trach should be changed only by ENT unless emergent, but this was not entered into the chart and was only communicated verbally and on a whiteboard. A second resident with anoxic brain damage, acute and chronic respiratory failure, tracheostomy status, and pneumonia due to pseudomonas experienced a fatal event related to a broken trach flange and trach displacement. A late-entry RT note documented that during first rounds at night, the RT found the trach tube broken and dislodged from the flange, with the tube about 1.5 inches out of position and deviated from midline. The RT called a nurse and another RT to assist, removed the old tube, and inserted a new Shiley XLT 6 that had been tested and lubricated. During suctioning, a significant amount of blood was observed, and repeated suction passes continued to remove blood. The SpO2 alarm indicated low oxygen saturation, the resident was disconnected from the ventilator for manual bagging, and SpO2 did not improve. A pulse check revealed no pulse, the resident was transferred to the floor, a smaller XLT 5 trach was placed without resistance, and manual ventilation and chest compressions continued until EMS arrived, but the resident could not be revived. A nursing late entry confirmed that the RT showed the nurse the broken trach, that vital signs were stable before the change, and that copious bleeding occurred during the trach change, followed by code blue and unsuccessful resuscitation. RT 4 reported that the trach was broken at the flange and deviated at a 45-degree angle, that the ventilator was not alarming despite the deviation, that the resident’s lips appeared cyanotic with SpO2 below 90%, and that he changed the trach to restore a patent airway. He described large amounts of frank blood and mucus on suctioning and believed the airway was filled with blood. RT 4 also reported concerns about ventilator alarms, pulse oximeters, alarm cables, and lack of needed supplies prior to the current RT Director’s tenure, and stated he had asked repeatedly for alarms, pulse oximeters, and cables. A third resident with quadriplegia, anoxic brain damage, and chronic respiratory failure with hypoxia and hypercapnia had orders related to BiPAP/ventilator use, including checking body positioning, nasal cannula placement, and BiPAP vent settings with humidity. The overall deficiency cited that staff did not follow physician orders for changing tracheostomies, did not obtain needed consultations with outside providers, did not accurately document resident condition prior to trach changes, and did not initiate hospital transfer when there were concerns about trach integrity. Staff interviews revealed knowledge of faulty trach equipment, including broken flanges that caused trach dislodgement, and that one resident with a broken flange and dislodged trach went into cardiac arrest and died. Additionally, another resident with displaced BiPAP was found unresponsive, CPR was initiated, and the resident could not be resuscitated and died. These failures in respiratory and tracheostomy care for multiple residents resulted in findings of immediate jeopardy for two of the residents.

Penalty

Fine: $27,471
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0695 citations
Unsecured Storage of Full Oxygen Cylinders on Nursing Unit
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Surveyors observed four full O2 cylinders on one nursing unit stored unsecured directly on the floor under a sign labeled "FULL CYLINDERS" instead of in a secured storage rack. The ADON confirmed the cylinders were full and should not be on the ground. Reference to NFPA 99 showed that freestanding cylinders must be protected from damage and properly chained or supported in a stand or cart. The DON and Maintenance Director both acknowledged that O2 cylinders are required to be stored in a secure rack, should never be on the floor, and that unsecured cylinders on the floor present a safety risk.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Improper Handling and Storage of Oxygen Nasal Cannula
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with severe cognitive impairment and COPD, receiving oxygen therapy via nasal cannula, was observed twice with the cannula lying on the floor beside the bed instead of stored in the bag on the oxygen concentrator as required. A CNA later picked up the cannula from the floor, wiped it with a non-disinfectant incontinent wipe, and reapplied it to the resident, despite having been trained that a cannula found on the floor should be replaced. An LVN, the DON, and the Administrator all confirmed that oxygen cannulas must be stored properly, replaced if found on the floor, and that incontinent wipes are not disinfectants, indicating a failure to follow the facility’s infection prevention and control policy.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Improper Storage of Nebulizer Mask and Respiratory Supplies
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with COPD and dementia, receiving scheduled nebulizer treatments, was found on multiple occasions to have a nebulizer mask stored on top of the machine rather than in a sanitary manner. A CNA and a nurse aide in training confirmed the mask’s placement, and an LPN reported that masks were routinely cleaned, dried, and then stored on top of the machine. The DON later acknowledged that masks should be washed, dried, and placed on a clean surface, and facility policy required oxygen and respiratory supplies to be stored in a plastic bag when not in use.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Oxygen Administered Without Required Physician Order
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident with acute respiratory failure with hypoxia, pulmonary hypertension, and type 2 diabetes was observed receiving oxygen at 4.5 L/min via nasal cannula without a corresponding physician order in the clinical record. The DON acknowledged that an order should have been in place before oxygen was initiated. Facility policy on supplemental oxygen via nasal cannula requires administration only under a physician or provider order, in alignment with 410 IAC 16.2-3.1-47(a)(6).

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Missing Current Physician Order for Oxygen
D
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

A resident was observed receiving O2 via nasal cannula on multiple occasions, but the chart had no current physician order for O2. The resident said she had been told after a recent hospitalization to use O2 for 30 days, but that time had passed and she was still using it because staff told her she needed it. The DON confirmed there was no current O2 order; the last order had already been discontinued.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Provide and Document Respiratory Care
E
F0695 F695: Provide safe and appropriate respiratory care for a resident when needed.
Short Summary

Failure to provide and document respiratory care: A resident with a trach had no documented evidence of respiratory rate, depth, and quality being monitored each shift and as needed, despite oxygen orders and trach care needs. Other residents with CPAP, nebulizer, and oxygen therapy had respiratory equipment left out of required storage, missing CPAP settings and care details in orders and care plans, and MAR entries signed by nursing staff even when respiratory staff reportedly completed the equipment changes.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Know what gets cited — and walk into your next survey with full visibility

We process and analyze inspection reports and Plans of Correction using AI to surface insights and trends — so you can improve care quality and stay ahead of compliance risk before your next survey.

Get ready for your next survey

See what surveyors are citing in your state and spot your risk areas before they do.

Monthly Citation Reports

Have you been cited for this tag?

Save hours drafting a compliant Plan of Correction — AI built on real approved POCs.

Plan of Correction Writer

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release June 24, 2026) and official state health department websites — never guesswork.

Trusted by long-term care providers and associations.

Allegria Senior Living logo
FHCA logo
WeCare Centers logo
Care Rehab logo
An unhandled error has occurred. Reload 🗙