F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
D

Incompatible Mattress and Bedframe for Dependent Resident

Savoy Care CenterMamou, Louisiana Survey Completed on 03-26-2026

Summary

The facility failed to ensure that a resident’s mattress was compatible with and properly fit the bedframe, as required by the expectation that all bed frames, mattresses, and bed rails be regularly inspected for safety and that mattresses attach safely to the bed frame. The affected resident had an admission date of 12/07/2023 and diagnoses including Myoneural Disorder, Paraplegia, Epilepsy, and Peripheral Vascular Disease, with a Quarterly MDS BIMS score of 6 indicating severe cognitive impairment. On 03/23/2026 at 11:39 a.m., surveyors observed the resident lying in bed with his feet hanging off the mattress, which appeared too small for the bedframe. On 03/24/2026 at 10:12 a.m., further observation showed the resident lying on his back with his head elevated on an air mattress that did not fit the bedframe properly, leaving approximately a 1-foot gap between the top of the bedframe and the head of the mattress, with the resident’s head partially above the mattress. At 12:48 p.m., a CNA confirmed the mattress did not fit the bedframe and explained that pulling the resident and mattress up in the bed would create a gap at the footboard. At 1:00 p.m., the ADON confirmed the mattress was not accommodating to the resident’s height and should be. At 3:00 p.m., the DON observed that the air mattress was approximately 1 foot smaller than the bedframe and confirmed that the bed was not accommodating the resident and that the mattress did not fit the bedframe properly.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Resources

Below are regulatory guidelines relevant to this citation:

See other F0909 citations
Failure to Inspect and Assess Bed Rails and Bed Systems for Entrapment Risk
D
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

Surveyors found that the facility did not conduct or document regular inspections or entrapment assessments of bed frames, mattresses, and bed rails, despite FDA guidance and manufacturer instructions requiring ongoing evaluation of hospital bed systems. Three residents with CVA, hemiplegia, cognitive impairment, and significant ADL dependence had quarter or half side rails in use, but therapy notes, care plans, and MDS assessments did not consistently reflect side rail use or any entrapment evaluations. Staff interviews revealed that the maintenance engineer only addressed side rails when reported broken, a RN was unaware of any side rail safety or inspection procedures, and a PT described routine admission of brain-injured residents with side rails and lap belts while indicating that nursing was responsible for bed safety zoning. The facility’s restraint policy addressed assessment and consent for positioning and safety devices but did not include any schedule or process for device inspection.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Missing Bed Entrapment Assessment and Incomplete Bed Inspection Logs
E
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

The facility failed to complete a bed entrapment assessment before a resident used bilateral upper bed rails and failed to document monthly inspections for all beds in use. A resident who had capacity to make decisions and needed partial/moderate assistance with mobility was observed using elevated side rails, and RN and Maintenance staff confirmed the rails were in use and that maintenance was responsible for the zone assessment. The February bed safety checklist did not show the required entrapment measurement for that resident, and the monthly bed maintenance log listed only a few rooms rather than all resident beds.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Loose Bed Rail Not Securely Attached
D
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

A resident with MS, seizures, generalized weakness, and impaired mobility had a loose right-hand mid-bed siderail that would not stay in place. Staff observed the rail drop to the floor when lifted, and a CMT confirmed it was loose and posed an injury risk. The Maintenance Director later found it attached with only one bolt near the head of the bed, and the facility could not provide documentation of weekly bed inspections.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Bed Rail Inspection and Compatibility Deficiency
D
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

Bed Rail Inspection and Compatibility Deficiency: The facility did not ensure regular entrapment inspections of bed systems for two residents using bilateral bed assist bars. Staff used a bed system measurement device that was missing a required scale component, did not retain test documentation, and were not following the bed manufacturer's specifications for compatible replacement rails. One resident had dementia, osteoporosis, and muscle weakness with impaired cognition, and another resident had MS and sepsis with substantial assistance needs for bed mobility.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Inspect Bed Side Rails
D
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

The facility failed to regularly inspect bed frames, mattresses, and side rails for four residents with diagnoses including fracture, COPD, autism, seizures, dementia, falls, spinal stenosis, MS, heart failure, and acute respiratory failure with hypoxia. Observations showed both half side rails upright and moving with minimal effort, and the medical records showed no maintenance inspection for the side rails. The DON said maintenance had not completed any inspections, the maintenance staff said the rails were placed when nursing requested and had not been inspected, and the Administrator expected side rail inspections at least quarterly.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Regularly Inspect Bed Rails and Bed Frames
D
F0909 F909: Regularly inspect all bed frames, mattresses, and bed rails (if any) for safety; and all bed rails and mattresses must attach safely to the bed frame.
Short Summary

Regular inspection of bed frames, mattresses, and bed rails was not conducted for two residents with bilateral upper quarter bed rails in use. One resident was observed in bed with the rails up, another stated she uses the rails if she needs to hold onto them and that staff use them, and the Maintenance Supervisor said staff only tighten loose bolts and do not keep logs of the work.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Know what gets cited — and walk into your next survey with full visibility

We process and analyze inspection reports and Plans of Correction using AI to surface insights and trends — so you can improve care quality and stay ahead of compliance risk before your next survey.

Get ready for your next survey

See what surveyors are citing in your state and spot your risk areas before they do.

Monthly Citation Reports

Have you been cited for this tag?

Save hours drafting a compliant Plan of Correction — AI built on real approved POCs.

Plan of Correction Writer

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release June 24, 2026) and official state health department websites — never guesswork.

Trusted by long-term care providers and associations.

Allegria Senior Living logo
FHCA logo
WeCare Centers logo
Care Rehab logo
An unhandled error has occurred. Reload 🗙