K0921 K921: Ensure that testing and maintenance of electrical equipment is performed.
D

Outdated Certification of Oxygen Concentrator

Spring Lake Rehabilitation CenterWinter Haven, Florida Survey Completed on 02-11-2025

Summary

The facility failed to maintain patient care related electrical equipment (PCREE) in accordance with NFPA 99 standards. During a facility tour, it was observed that an oxygen concentrator in the rehabilitation area had an outdated PCREE certification, with the last certification dated July 2021. This indicates that the equipment had not been tested or certified for compliance for a significant period, contrary to the requirements that all PCREE be tested before being put into service and after any repair or modification. The maintenance director, who was present during the tour, acknowledged that the concentrator was provided by a rental company. This suggests a lapse in the facility's oversight of equipment maintenance and certification, particularly for equipment sourced externally. The findings were discussed with the maintenance director and the administrator during the exit conference, highlighting the facility's failure to adhere to established protocols for electrical equipment testing and maintenance.

Plan Of Correction

1. Identified concentrator was removed from service, PCREE certification service completed. 2. Maintenance Director/Designee completed observation of other facility concentrators to verify PCREE certification is completed as required. 3. The administrator provided education for current facility Maintenance Department regarding PCREE maintenance requirements. 4. Maintenance Director/Designee to complete monthly observation of facility concentrators to ensure PCREE certification is completed as required. Findings to be reviewed at the monthly QAPI Committee Meeting. Modifications as indicated based on findings.

Penalty

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
See other K0921 citations
Failure to Maintain and Inspect Patient-Care Electrical Equipment
E
K0921 K921: Ensure that testing and maintenance of electrical equipment is performed.
Short Summary

Surveyors found that fixed patient-care electrical equipment was not properly maintained or inspected in accordance with NFPA 99. In one room, a bedside remote had mismatched insulation and exposed wiring, and in another room, a call button receptacle had exposed low-voltage conductors. The Maintenance Director acknowledged both issues and reported that new bed remotes had been received but not yet installed.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Maintain Annual Testing Documentation for Patient-Care Electrical Equipment
F
K0921 K921: Ensure that testing and maintenance of electrical equipment is performed.
Short Summary

Surveyors found that the facility did not maintain required documentation showing that annual NFPA 99-compliant testing (including physical integrity, resistance, leakage current, and touch current tests) was performed on electrical equipment used for patient care. During record review with the Maintenance Director, no records could be produced to verify that this testing had been completed as required, and the Maintenance Director acknowledged the lack of documentation.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Maintain and Inspect Patient-Care Electrical Equipment per NFPA 99
D
K0921 K921: Ensure that testing and maintenance of electrical equipment is performed.
Short Summary

Surveyors found that patient-care related electrical equipment was not inspected and maintained in accordance with NFPA 99 requirements. Biomedical inspection records showed that inspections were not current, and an observation in the gym revealed two microwaves, a toaster oven, and an "E Stem" device with overdue inspection stickers from 2024. During an interview, an assistant reported that this equipment was supplied by the corporate office and maintained by a different company, yet the items remained in use in a patient-care area without up-to-date inspections, affecting one of seven smoke compartments.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Test Electrical Equipment Before Use
E
K0921 K921: Ensure that testing and maintenance of electrical equipment is performed.
Short Summary

A portable blood pressure machine in the facility was found to be in use without undergoing the required Bio-Medical testing. The machine had been used daily for over two months, as confirmed by a staff nurse and the Director of Maintenance. This deficiency was acknowledged by the facility's administration during the exit conference.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Perform Annual Electrical Inspections
F
K0921 K921: Ensure that testing and maintenance of electrical equipment is performed.
Short Summary

The facility did not conduct the required annual inspection of electrical receptacles in resident care areas, affecting all resident bed locations. This deficiency was confirmed during an exit interview with the Administrator, Director of Maintenance, and Assistant.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.
Failure to Maintain Electric Beds as per Manufacturer's Guidelines
D
K0921 K921: Ensure that testing and maintenance of electrical equipment is performed.
Short Summary

The facility did not ensure proper maintenance of electric beds as per the manufacturer's guidelines. During a survey, it was found that electrically operated beds were not inspected at the required intervals, and there was no documentation of formal inspections or maintenance. The Environmental Services Director acknowledged the lack of unique identifiers for beds and stated that beds were checked annually using a room audit form, but specific beds were not identified. This failure to adhere to maintenance requirements constituted a deficiency in compliance with NFPA 99 standards.

No penalty information released
tooltip icon
The penalty, as released by CMS, applies to the entire inspection this citation is part of, covering all citations and f-tags issued, not just this specific f-tag. For the complete original report, please refer to the 'Details' section.

Know what gets cited — and walk into your next survey with full visibility

We process and analyze inspection reports and Plans of Correction using AI to surface insights and trends — so you can improve care quality and stay ahead of compliance risk before your next survey.

Get ready for your next survey

See what surveyors are citing in your state and spot your risk areas before they do.

Monthly Citation Reports

Have you been cited for this tag?

Save hours drafting a compliant Plan of Correction — AI built on real approved POCs.

Plan of Correction Writer

Trusted data from CMS and state health departments

Every citation, penalty and Plan of Correction is sourced from public CMS records (latest release June 24, 2026) and official state health department websites — never guesswork.

Trusted by long-term care providers and associations.

Allegria Senior Living logo
FHCA logo
WeCare Centers logo
Care Rehab logo
An unhandled error has occurred. Reload 🗙